Recordings of Impedance and Communication Between Defibrillator and Pacemaker Electrodes

Recordings of Impedance and Communication Between Defibrillator and Pacemaker Electrodes

Anders Jarløv, Anne Elisabeth Jarløv, Tim Toftgaard Jensen
Copyright: © 2019 |Pages: 24
DOI: 10.4018/IJBCE.2019070103
OnDemand:
(Individual Articles)
Available
$37.50
No Current Special Offers
TOTAL SAVINGS: $37.50

Abstract

External defibrillation involves short electric shocks of several thousand volts applied to the chest of patients suffering from heart stop. The shock coordinates disorganized muscle fibers of the heart which then can resume normal activity. Implanted pacemakers monitor the natural electrical activity of the heart via electrodes mounted permanently on the inside of the heart. If this activity is insufficient, the pacemaker stimulates the heart muscle by applying a short electrical impulse of a few volts. During the period of defibrillation, voltage differences arise simultaneously between various locations of the body preventing recordings of the heart's natural activity. In order to quantify parameters determining voltage and current in a conductive medium or in a myocardium when defibrillator and pacemaker electrodes are present simultaneously, impedances between electrodes positioned on conductive materials were recorded in laboratory set ups, and the methods were tested using a porcine heart in vitro.
Article Preview
Top

Introduction

External defibrillation is performed in order to resuscitate patients suffering from cardiac arrest by applying an electric shock of several thousand volts to the patient’s thorax for a number of milliseconds giving rise to currents in the order of several amperes passing through the tissues of the thorax and through the myocardium. This throws the heart into refractoriness for a few seconds initiating normal myocardial function according to e.g. Sperilakis, N. (1989) and Hall, J. E. (2016). Defibrillation of the exposed heart is in principle performed in the same way using in the order of hundreds of volts according to Schwarz B et al. (2003). External and internal defibrillation are also used to eliminate arrhythmia of the heart (Thakor, N. V., 1984). Internal defibrillation involves the application of cardiac catheters with electrodes in the distal end. These are applied to the inside of the atrium and electric stimuli of different time courses and strength are used as described by Cooper, R.A.S., Johnson, E.E. and Wharton, J. M. (1997).

External defibrillation, involving energy levels up to 400 Joules, may cause tissue damage in the form of burn marks to the skin (Patriciu, A. et.al.2005) and injuries to underlying tissues such as imbalance of the ion concentration inside the cells and in the body fluids e. g. the interstitial and intercellular fluids dealt with by Patel & Roth (2001) and Kim, Y., Zieber, H. G. and Wang, I. E. (1990). This may have an unfavorable influence on the electrical potentials between the inside and the immediate outside of the cardiac nerve fibers which normally are of the order of -80 to -95 mV (Sperilakis, N., 1989). Biphasic and damped sine wave shocks are dealt with by Schwarz, B. et al (2003). Internal atrial defibrillation is dealt with by Cooper, R. A. S., Johnson, E. E., & Wharton, J. M., (1997) using monophasic and biphasic stimulation and by Patel, S. G. & Roth, B. J. (2001) demonstrating polarization of tissue in the vicinity of surface electrodes in an electric field. Defibrillation of the exposed heart using a single mono phasic shock has been chosen as a scenario for the experiments to be accounted for in the following.

Pacemakers are electric stimulators that can monitor electric activity (ECG) of the heart and intervene in cases of arrhythmia by supplying electric stimuli in the order of a few volts and of a duration of a half to one millisecond to the heart muscle (Escher, D., 1973). Implanted pacemaker electrodes are generally fastened to the inside of the heart ventricle (Starr, D. et al, 1981, Reade, M. C., 2007; and Schaldach, M., 1971). Flexible cords connect the electrodes to the pacemaker itself, which is placed subcutaneously in the chest of the patient. Implanted pacemakers can be controlled from the outside regarding “On- Off” as well as amplitude and frequency of stimulus and can work for a number of years.

To determine the correlation between defibrillation voltages and currents in conductive media and the influence of these on the voltage between implanted bipolar pacemaker electrodes, recordings of voltage and impedance between sets of stylized electrodes placed on conductive media was performed in the frequency range from 25cps to 2000cps. The results were compared to results retrieved from an empirically derived formula (Jarlov, A. & Jensen, T. T., 2017) determining the electrical impedance between the electrodes. Defibrillator shocks as well as pacemaker stimuli were afterwards applied to a porcine myocardium in vitro and the electrode voltages were recorded by a four-channel oscilloscope. The experiments and procedures described in the following are meant to illustrate measuring principles and the application of analogue models of the electrical circuits used in the experiments and are in no way recommended for clinical use. The designations “Pacemaker” and “Defibrillator” are used freely to cover commercially available instruments as well as laboratory set ups used as their substitutes.

Complete Article List

Search this Journal:
Reset
Volume 12: 1 Issue (2024): Forthcoming, Available for Pre-Order
Volume 11: 2 Issues (2022): 1 Released, 1 Forthcoming
Volume 10: 2 Issues (2021)
Volume 9: 2 Issues (2020)
Volume 8: 2 Issues (2019)
Volume 7: 2 Issues (2018)
Volume 6: 2 Issues (2017)
Volume 5: 2 Issues (2016)
Volume 4: 2 Issues (2015)
Volume 3: 2 Issues (2014)
Volume 2: 2 Issues (2013)
Volume 1: 2 Issues (2012)
View Complete Journal Contents Listing